Jim,

I checked the two plots you made. Yes, they
give us a strong indication that the resolution
of |tk_tg-tk| is correlated with ek_tg. i will 
take this into account in modification of TGTCON
algorithm. thanks.

for the plots about TARGF cut i sent to you yesterday,
i made them using the old km21 ntuples (generated around
May 5-7), which means your recent update of CCD pulse fit 
and some new categories in TG reconstruction are not there. 
So, when Joe has the new ntuples ready, i will check the same
events and see if any improvements could be achieved.

for CCD pulse fitting, in my study i found that the result
are still not that reliable even we require chi(1)/chi(2)
greater than 2.5- we have to put a cut on the fitted area 
for the second pulse. i don't know if you have already 
considered this point in your target pulse fitting.

jingliang

> Another question, and a point of interest:
> 
> 1) The TGTCON cut is set at 3.75ns.  I think
> the cut should be made as a function of ek_tg.
> For example, the resolution for 0.5 MeV kaon
> fibers is very much worse than for 5 MeV fibers.
> Though you're losing little acceptance by this
> cut, maybe there's a more efficient way to
> make the cut.
> 
> A quick look at some plots on bnlku36:
> /home/scratch/frank/target/p2nt/ might illustrate
> what I mean.  Using a few runs of km21 monitors,
> tgtcon.ps shows the energy distribution of a
> sample of ek_tg.  The top show those with
> |tk_tg-tk|<2, the overlay 2<|tk_tg-tk|<4.
> This shows a 'heavy' component of the time
> outliers have energy<3 MeV.  The plot
> tgtcon2.ps shows the tk_tg-tk for 3
> different ek_tg bins: 10-30MeV, 3-10MeV,
> and .lt.3 MeV.  Again, the correlation
> is obvious.
> 
> 2)  Since the target pulse fitting is (hopefully) working, 
> I've been looking at the problem with the 2nd pulse energy
> scale.  (I'm convinced the scale is wrong using the
> pulsate algorithm).  I think that we will be able to
> gain some acceptance if we 'correct' the epi_tg using
> the fitted energy, rather than the 'pulsate' energy.
> This will primarily be the epimax type of cuts, as
> well as the opposite side pion cut.  Maybe 1% is
> realizable.
> 
> Jim
> 
> 